Only one set of tag calibration settings is allowed per tag. Therefore you should use the tag calibration settings to define tag-specific data such as alarm / trip points, input and output ranges, and units of measure. Other calibration values that are not tag-specific such as number of calibration points, tolerance, and so forth, can be stored in the calibration profile settings.

  1. Start the Calibration module.

  2. Select Actions > Tag Settings .

  3. On the Enter Tag Number dialog, type the required tag number or select Find to search instrument tags.

    If you selected multiple tag numbers, select Actions > Tag List or PPM All Outputs Graphic and then select a tag number to define the settings for that instrument.

  4. Select Actions > Select Profile

  5. Define the calibration settings values for the tags as required .

    • If you are defining calibration settings for a tag number associated with process data, you can copy existing process data minimum and maximum range values. For details, see Copy Existing Process Data.

    • If you are using the Standard Settings calibration sheet, the Measuring Point parameter is available for single point calibrations only.

    • If you are using the Standard Settings calibration sheet, the fields in the Set values group box are available only if under Calibration type, you selected Switch set point.

  6. Select Save Save Icon 16bit to save the settings.